product-template-default,single,single-product,postid-270,qode-quick-links-1.0,woocommerce,woocommerce-page,ajax_fade,page_not_loaded,, vertical_menu_transparency vertical_menu_transparency_on,qode-title-hidden,paspartu_enabled,qode_grid_1300,columns-4,qode-product-single-tabs-on-bottom,qode-theme-ver-11.2,qode-theme-bridge,wpb-js-composer js-comp-ver-5.2.1,vc_responsive

UV-3600

The UV-3600 UV-VIS-NIR Spectrophotometer is a new additional model in our product lineup of high-end UV-VIS-NIR spectrophotometers.
The UV-3600, a successor model to the UV-3150/3101, is designed for the measurement of liquid samples, while the SolidSpec-3700/3700DUV is mainly designed for the measurement of solid samples.

Features

 

High Sensitivity – The Only 3 Detector System
Equipped with 3 detectors: A PMT detector (photomultiplier tube) for the ultraviolet and visible regions, and InGaAs and PbS detectors for near infrared region. The InGaAs detector bridges the gap between the PMT – PbS switching wavelength where sensitivity is typically low to ensure high sensitivity over the entire measurement wavelength range. With <0.00003 Abs noise at 1500 nm, the world´s lowest noise level is achieved.

 

High Resolution – Wide Measurement Range & Ultra-low Stray Light
With its high performance double monochromator, ultra-low stray light (0.00005% or less at 340 nm) is achieved at high resolution (highest resolution 0.1 nm). With a measurement wavelength range of 185 – 3300 nm, measurement is possible over a wide range including the ultraviolet, visible and near infrared regions. This allows spectroscopic analysis in a wide variety of fields.

 

Software
All-in-one Software
UVProbe software for the UV-3600 includes functions for spectrum measurement, quantitation, kinetics and a report generator. Multiple security and audit-trail functions ensure the reliability of data.

SPECIFICATION

UV-3600
Wavelength range 185nm – 3300nm
Spectral Bandwidth 8 steps in ultraviolet/visible region 0.1, 0.2, 0.5, 1, 2, 3, 5, 8 nm
10 steps in near-infrared region 0.2, 0.5, 1, 2, 3, 5, 8, 12, 20, 32 nm
Resolution 0.1 nm
Sampling Pitch 0.01 – 5 nm
Wavelength accuracy UV/VIS region: ±0.2 nm NIR region: ±0. 8 nm
Wavelength repeatability UV/VIS region: less than ±0.08 nm NIR region: less than ±0.32 nm
Stray light < 0.00008% (220 nm, NaI)
< 0.00005% (340 nm, NaNO2)
< 0.0005% (1420 nm, H2O)
< 0.005% (2365 nm, CHCl3)
Photometric range -6 to 6 Abs
Noise 0.00005 Abs or less (500 nm), 0.00008 Abs or less (900 nm), 0.00003 Abs or less (1500 nm)
Slit width 2nm, RMS value at 1 sec. response
Photometric System Double beam
Dimensions 1020(W) x 660(D) x 275(H) mm